ORIGINAL ARTICLE
Year : 2018  |  Volume : 8  |  Issue : 5  |  Page : 446-450
Biomechanical behavior of an implant system using polyether ether ketone bar: Finite element analysis


São Leopoldo Mandic Dental Research Center, Campinas, SP, Brazil

Date of Submission14-May-2018
Date of Acceptance11-Jul-2018
Date of Web Publication8-Oct-2018

Correspondence Address:
Prof. Simone Kreve
Rua Independencia, 1899, Apto 602, Centro, Toledo, Parana
Brazil
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jispcd.JISPCD_183_18

Rights and Permissions

   Abstract 

Aim and Objectives: This study assessed, through finite element analysis, the biomechanical behavior of an implant system using the All-on-Four® technique with nickel–chromium (M1) and polyether ether ketone (PEEK) bars (M2).
Materials and Methods: Implants and components were represented in three-dimensional (3D) geometric models and submitted to three types of load: axial, oblique, and load on all teeth. The 3D models were exported to a computer-aided design-like software such as Solidworks 2016 (Dassault Systemes, Solidworks Corps, USA) for editing and Nonuniform Rational Basis Splines parametrization.
Results: Data were analyzed according to system's areas of action: peri-implant bone, implant, intermediates, intermediates' screws, prostheses' screws, and bars. Largest peak stress was shown in M2.
Conclusion: PEEK is a promising material for use in dentistry; however, further studies are necessary to evaluate its performance.


Keywords: Biomedical and dental materials, dental implant-abutment design, dental implants, finite element analysis


How to cite this article:
Jaros OA, De Carvalho GA, Franco AB, Kreve S, Lopes PA, Dias SC. Biomechanical behavior of an implant system using polyether ether ketone bar: Finite element analysis. J Int Soc Prevent Communit Dent 2018;8:446-50

How to cite this URL:
Jaros OA, De Carvalho GA, Franco AB, Kreve S, Lopes PA, Dias SC. Biomechanical behavior of an implant system using polyether ether ketone bar: Finite element analysis. J Int Soc Prevent Communit Dent [serial online] 2018 [cited 2019 Dec 6];8:446-50. Available from: http://www.jispcd.org/text.asp?2018/8/5/446/242861



   Introduction Top


The loss of dental elements has a profound effect on people's quality of life. The use of implants comes as an important step for rehabilitation.[1],[2]

In some cases, there are anatomical limitations for implant placement.[3] One of the techniques developed to circumvent these limitations is the placement of tilted implants on regions close to the maxillary sinus wall and lower alveolar nerve.[4],[5]

In 2003, the concept All-on-Four® was introduced, consisting of two parallel anterior implants and two tilted posterior implants that fixate a total prosthesis.[6] These distal implants' placement helps to reduce the prosthesis cantilever,[7] thus reducing stress created during mastication,[4],[8],[9] which is transferred to the prosthesis, implant, and bone structures. This stress can influence bone remodeling.[10],[11]

New technologies have been developed to fabricate accurate metal implant-supported frameworks.[12]

Finite element model is a computer method for stress distribution analysis that has been used for creating virtual model.[13] The effect of loading strengths over dental implant elements and peri-implant bone can be recorded by applying the equivalent stress (von Mises stress) expressed in Megapascals (MPa).[13],[14] The computational models also enable the analysis of the stress distribution occurring on the prosthesis-implant system and the load variations in different system designs.[11]

It is important to identify the degree of misfit between implant-supported frameworks and implants/abutments that commonly occur in implant dentistry. Computer-aided design (CAD)/computer-aided manufacturing scanning technology was developed to be a viable method to measure the amount of misfit between implant-supported frameworks and implant restorative components.[15],[16]

Polyether ether ketone (PEEK) is a biomaterial currently applied to dental and medical devices[17],[18] that has been studied as an alternative to metal alloys in protocol type bars.[19] It is a linear high-performance polymer due to its mechanical properties, stability at high temperatures, and chemical resistance. PEEK also shows coloration similar to that of the teeth, low weight and is an alternative to patients with metal allergies.[20]

In this context, it is important to study the biomechanical behavior of peri-implant bone, implants, and prosthetic components of nickel–chromium (Ni-Cr) and PEEK All-on-Four® protocol, under physiologic occlusal loads.


   Materials and Methods Top


This work was approved by the Ethics Committee (2016/0708) of São Leopoldo Mandic Dental Research Center.

For the virtual models, we scanned a total lower prosthesis and a jawbone (Nacional Ossos, Jaú, São Paulo, Brazil) using a laser three-dimensional (3D) scanner (Nextengine HD, Santa Monica, USA). For a better scan, we used commercial talc (Talco Baby, Johnson e Johnson, New Brunswick, Nova Jersey, USA) on the jawbone and sprayed matte acrylic white paint (Suvinil, Basf Brasil SA, São Paulo, Brazil) on the total prostheses to avoid laser reflection and potential distortions during scanning.

Prosthesis and jawbone were circularly scanned 16 times at intervals of 22.5°. The models were then recorded in SLT format (3D Systems, Rock Hill, USA) for processing.

The 3D models were exported to a CAD-like software Solidworks 2016 (Dassault Systemes, Solidworks Corps, USA) for editing and Nonuniform Rational Basis Splines parametrization.

The cortical and medullary bone representation was 2.0 mm thick and was considered Type III bone.[21] The implant-supported total prosthesis outer geometry was represented by the delimitation of the acrylic base and teeth. Muscle insertions on the jawbone were also simulated to stabilize the system when load is applied. Implants and prosthetic components' representations were supplied by the manufacturer (SIN-Sistemas de Implantes, São Paulo, Brazil).

Combining the scanned models and computational models, we obtained an implant-supported fixed total prosthesis with the following characteristics:

  • Implants measuring 3.75 mm × 13 mm (Strong Sw Hexagonal Extern, SIN – Sistemas de implantes, São Paulo, Brazil). The platform was placed at the crest level with two anterior implants parallel and perpendicular to the bone crest and two posterior implants tilted in 30° relative to the long axis of the anterior implants and at 3 mm anteriorly to the mental foramen
  • Intermediates with 4 mm of height, anterior implants upright, and posterior ones tilted in 30°
  • Titanium screws
  • PEEK and Ni-Cr bars measuring 3.5 mm × 5 mm, with round corners and cantilever of 15 mm
  • Acrylic gingiva and acrylic resin stock teeth
  • Saucerization of 1.5 mm.[22],[23]


Masticatory load simulated on teeth 14, 15, and 16 with three contact points measuring 1.0 mm of diameter on each element and single point load on all teeth. Food bolus was modeled with 5 mm thick.

We designed two different models for the simulation:

  • Model M1 (control): Prosthesis with Ni-Cr bar
  • Model M2: Prosthesis with PEEK bar.


The models were exported from Solidworks to the finite element simulation software Ansys Workbench V17.2 (Ansys Inc., Canonsburg, PA, USA). The mechanical behavior of each component was set with their respective elasticity modulus and Poisson coefficient according to the literature.[24],[25] All implants were considered osseointegrated.

Simulation used masticatory physiological loads of 150 N as posterior unilateral loads. The load on all teeth was simulated as 60 N on molars, 40 N on premolars, and 20 N on anteriors.[26],[27]


   Results Top


Peri-implant bones were analyzed according to Mohr-Coulomb criterion. Calculation considered tensile yield strength of 82.8 MPa and compressive yield strength of 133.6 MPa [Figure 1] and [Figure 2].[28] Model M1 was set as control.
Figure 1: Peaks on peri-implant bone according to Mohr coulomb criterion (in megapascals)

Click here to view
Figure 2: Results for the peri-implant bone under oblique load (outer and sectional view)

Click here to view


Due to its high ductility, implants were analyzed according to von Mises criterion with yield of 550 MPa [Figure 3] and [Figure 4].
Figure 3: Peak stress on implants according to von Mises criterion (in megapascals)

Click here to view
Figure 4: Results of left posterior implant under oblique load (vestibular view [V])

Click here to view


Intermediate, intermediate's screws, and prostheses' screws were analyzed according to von Mises criterion and yield was considered 880 MPa [Figure 5].
Figure 5: (a) Results of the left posterior implant under load for all teeth (lingual view-L). (b) Results for the intermediate's screw and left posterior screw under oblique load (lingual view-L)

Click here to view


On all three cases, M2 showed the largest peak stress.

The bar was analyzed using Rankine method. Tensile yield strength was 524.7 MPa and 100 MPa for PEEK [Figure 6] and [Figure 7].
Figure 6: Percent of tensile yield strength on the outer sides peaks of the bars according to Rankine criterion (in megapascals)

Click here to view
Figure 7: Maximum main stress under load on all teeth on nickel–chromium and polyether ether ketone bars (apical view-A)

Click here to view



   Discussion Top


Finite element method has been widely used in dental and medical research to assess the simulated distribution of forces, which is virtually impossible in a clinical setting.[7],[26]

Results obtained on the peri-implant bone show that, for both axial and oblique loads, M2 suffered higher risk of bone loss compared with M1. According to Mohr-Coulomb criterion, this could suggest bone fracture since results were larger than 1.0 in both models; however, this could only lead to bone remodeling.[29],[30],[31] Studies have shown that hardness differences influence load distribution to the system.[11] When the all-teeth load was applied on the posterior implants, both systems behaved similarly; however, a significant difference was observed in the anterior implants, possibly due to the difference in hardness of the infrastructures, which helps to distribute stress, agreeing with another study.[27] When comparing these results with the force dissipation patterns on the bone associated with natural teeth, we notice that, under axial load, the peaks tend to occur on the furcation. Under oblique load, stress peaks occur on the vestibular cervical region.[32] In this study, stress was accumulated on the cervical, vestibular, and distal regions due to the posterior implants' angle and to the occlusal contacts located at the vestibular cusps.

On the implants under axial load, M1 peak occurred at the anterior and lingual region of the outer surface of the left posterior implants first thread, while M2 peak occurred at the lingual posterior region. Nonetheless, the difference was little significant on most implants. The fact that the bone is less rigid than the implants make it more easily deformable, minimizing deformity of more rigid structures.[24],[27] Under oblique load, implants showed a significant difference for model M2. Under all-teeth load, the implants' behavior was similar in both models.

The intermediates' screws also present a preload stress to simulate initial stress.[33] Peaks generated under oblique load were similar to those generated under axial load. The load on all teeth generated peaks on the screw's first thread, as well as on the other loads.

Prostheses screws under axial load sustained peaks on the same regions than those of the intermediates' screws, corroborating other studies.[27] The same was observed under oblique load. Under oblique load, M2 model was significantly different from M1 (67%). This can be explained by the high deformity on PEEK bar if compared to the Ni-Cr infrastructure deformity.

Under axial load, peaks occurred on the bars adjacent to the contact surface between the left posterior implant's intermediate and the bar. Given the difference in rigidity of Ni-Cr and PEEK, smaller values are expected on the same load conditions, as shown by other authors,[34],[35],[36] particularly by de Carvalho et al.,[19] who showed that metal bars present higher compression strength compared to PEEK bars, regardless of its design. Under oblique load, M1 model presented peaks at the lingual region of the left posterior implant cavum. In model M2, stress was concentrated adjacent to the contact surface between left posterior implant's intermediate and bar. Under oblique load, Ni-Cr bars show superior performance, although both materials have shown peaks far from the materials resistance, suggesting a favorable prognostic in both cases, as previously suggested.[37],[38] A smaller elasticity modulus offers less tensile strength, and as a consequence, smaller load on the components and smaller bar deformity. Under all-teeth load and the other loads, the infrastructure showed a significant difference between materials, but both are far from the materials' resistance, suggesting a long lifespan in clinical conditions, although M1 presents longer lifespan.


   Conclusion Top


Both under axial load and under load on all teeth, the two implant systems of the peri-implant bone showed similar stress distribution, and under oblique load, the PEEK bar system showed larger stress transference.

On implants, under axial and all-teeth loads, systems supported similar stress; under oblique load, PEEK bar system showed a larger stress peak.

Concerning prostheses and components' screws, the observed behavior was similar except for the PEEK models, which showed larger oblique load.

PEEK bar showed larger stress peaks in all simulations.

Although PEEK has been shown as a promising material, further studies are necessary to improve its usage and to enable its use as material for prostheses on implants.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Özdemir Doğan D, Polat NT, Polat S, Şeker E, Gül EB. Evaluation of “all-on-four” concept and alternative designs with 3D finite element analysis method. Clin Implant Dent Relat Res 2014;16:501-10.  Back to cited text no. 1
    
2.
Sargozaie N, Moeintaghavi A, Shojaie H. Comparing the quality of life of patients requesting dental implants before and after implant. Open Dent J 2017;11:485-91.  Back to cited text no. 2
    
3.
Elakkiya S, Ramesh AS, Prabhu K. Systematic analysis on the efficacy of bone enhancement methods used for success in dental implants. J Indian Prosthodont Soc 2017;17:219-25.  Back to cited text no. 3
[PUBMED]  [Full text]  
4.
Krekmanov L, Kahn M, Rangert B, Lindström H. Tilting of posterior mandibular and maxillary implants for improved prosthesis support. Int J Oral Maxillofac Implants 2000;15:405-14.  Back to cited text no. 4
    
5.
Calandriello R, Tomatis M. Simplified treatment of the atrophic posterior maxilla via immediate/early function and tilted implants: A prospective 1-year clinical study. Clin Implant Dent Relat Res 2005;7 Suppl 1:S1-12.  Back to cited text no. 5
    
6.
Maló P, de Araújo Nobre M, Lopes A, Ferro A, Gravito I. All-on-4® treatment concept for the rehabilitation of the completely edentulous mandible: A 7-year clinical and 5-year radiographic retrospective case series with risk assessment for implant failure and marginal bone level. Clin Implant Dent Relat Res 2015;17 Suppl 2:e531-41.  Back to cited text no. 6
    
7.
Sannino G. All-on-4 concept: A 3-dimensional finite element analysis. J Oral Implantol 2015;41:163-71.  Back to cited text no. 7
    
8.
Maló P, Rangert B, Nobre M. “All-on-four” immediate-function concept with Brånemark System implants for completely edentulous mandibles: A retrospective clinical study. Clin Implant Dent Relat Res 2003;5 Suppl 1:2-9.  Back to cited text no. 8
    
9.
Lopes A, Maló P, de Araújo Nobre M, Sanchez-Fernández E. The nobelGuide® all-on-4® treatment concept for rehabilitation of edentulous jaws: A prospective report on medium- and long-term outcomes. Clin Implant Dent Relat Res 2015;17 Suppl 2:e406-16.  Back to cited text no. 9
    
10.
Frost HM. Bone “mass” and the “mechanostat”: A proposal. Anat Rec 1987;219:1-9.  Back to cited text no. 10
    
11.
Rubo JH, Souza EA. Finite element analysis of stress in bone adjacent to dental implants. J Oral Implantol 2008;34:248-55.  Back to cited text no. 11
    
12.
Ortorp A, Jemt T. Clinical experience of CNC-milled titanium frameworks supported by implants in the edentulous jaw: A 3-year interim report. Clin Implant Dent Relat Res 2002;4:104-9.  Back to cited text no. 12
    
13.
Bramanti E, Cervino G, Lauritano F, Fiorillo L, D'Amico C, Sambataro S, et al. FEM and von mises analysis on prosthetic crowns structural elements: Evaluation of different applied materials. ScientificWorldJournal 2017;2017:1029574.  Back to cited text no. 13
    
14.
Cervino G, Romeo U, Lauritano F, Bramanti E, Fiorillo L, D'Amico C, et al. Fem and von mises analysis of OSSTEM® dental implant structural components: Evaluation of different direction dynamic loads. Open Dent J 2018;12:219-29.  Back to cited text no. 14
    
15.
Riedy SJ, Lang BR, Lang BE. Fit of implant frameworks fabricated by different techniques. J Prosthet Dent 1997;78:596-604.  Back to cited text no. 15
    
16.
Drago C, Saldarriaga RL, Domagala D, Almasri R. Volumetric determination of the amount of misfit in CAD/CAM and cast implant frameworks: A multicenter laboratory study. Int J Oral Maxillofac Implants 2010;25:920-9.  Back to cited text no. 16
    
17.
Kurtz SM, Devine JN. PEEK biomaterials in trauma, orthopedic, and spinal implants. Biomaterials 2007;28:4845-69.  Back to cited text no. 17
    
18.
Najeeb S, Zafar MS, Khurshid Z, Siddiqui F. Applications of polyetheretherketone (PEEK) in oral implantology and prosthodontics. J Prosthodont Res 2016;60:12-9.  Back to cited text no. 18
    
19.
de Carvalho GA, Franco AB, Kreve S, Ramos EV, Dias SC, do Amaral FL. Polyether ether ketone in protocol bars: Mechanical behavior of three designs. JIOH 2017;9:202-9.  Back to cited text no. 19
    
20.
Schwitalla AD, Spintig T, Kallage I, Müller WD. Flexural behavior of PEEK materials for dental application. Dent Mater 2015;31:1377-84.  Back to cited text no. 20
    
21.
Lekholm U, Zarb FL. Patient selection and preparation. Tissue integrated prostheses: Osseointegration in clinical dentistry. Quintessence 1985;54:611-2.  Back to cited text no. 21
    
22.
Wheeler SL. Implant complications in the esthetic zone. J Oral Maxillofac Surg 2007;65:93-102.  Back to cited text no. 22
    
23.
Browaeys H, Dierens M, Ruyffelaert C, Matthijs C, De Bruyn H, Vandeweghe S, et al. Ongoing crestal bone loss around implants subjected to computer-guided flapless surgery and immediate loading using the all-on-4® concept. Clin Implant Dent Relat Res 2015;17:831-43.  Back to cited text no. 23
    
24.
Holmes DC, Diaz-Arnold AM, Leary JM. Influence of post dimension on stress distribution in dentin. J Prosthet Dent 1996;75:140-7.  Back to cited text no. 24
    
25.
Overview of materials for Polyetheretherketone, Unreinforced 2017. Available from: Matweb.com. [Last accessed on 2017 Apr 15].  Back to cited text no. 25
    
26.
Bevilacqua M, Tealdo T, Pera F, Menini M, Mossolov A, Drago C, et al. Three-dimensional finite element analysis of load transmission using different implant inclinations and cantilever lengths. Int J Prosthodont 2008;21:539-42.  Back to cited text no. 26
    
27.
Menini M, Pesce P, Bevilacqua M, Pera F, Tealdo T, Barberis F, et al. Effect of framework in an implant-supported full-arch fixed prosthesis: 3D finite element analysis. Int J Prosthodont 2015;28:627-30.  Back to cited text no. 27
    
28.
Bayraktar HH, Morgan EF, Niebur GL, Morris GE, Wong EK, Keaveny TM, et al. Comparison of the elastic and yield properties of human femoral trabecular and cortical bone tissue. J Biomech 2004;37:27-35.  Back to cited text no. 28
    
29.
Hart RT, Davy DT, Heiple KG. A computational method for stress analysis of adaptive elastic materials with a view toward applications in strain-induced bone remodeling. J Biomech Eng 1984;106:342-50.  Back to cited text no. 29
    
30.
Natali AN, Pavan PG, Ruggero AL. Analysis of bone-implant interaction phenomena by using a numerical approach. Clin Oral Implants Res 2006;17:67-74.  Back to cited text no. 30
    
31.
Taylor D, Hazenberg JG, Lee TC. Living with cracks: Damage and repair in human bone. Nat Mater 2007;6:263-8.  Back to cited text no. 31
    
32.
Franco AB, Carvalho GA, Dias SC, Kreve S, De Martin AS. Osteoporosis and endodontic access: Analysis of fracture using finite element method. IJODM 2017;16:1-5.  Back to cited text no. 32
    
33.
Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-36.  Back to cited text no. 33
    
34.
Goll GE. Production of accurately fitting full-arch implant frameworks: Part I – Clinical procedures. J Prosthet Dent 1991;66:377-84.  Back to cited text no. 34
    
35.
Akça K, Cehreli MC, Iplikçioglu H. A comparison of three-dimensional finite element stress analysis with in vitro strain gauge measurements on dental implants. Int J Prosthodont 2002;15:115-21.  Back to cited text no. 35
    
36.
Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JY. Clinical complications with implants and implant prostheses. J Prosthet Dent 2003;90:121-32.  Back to cited text no. 36
    
37.
Benzing UR, Gall H, Weber H. Biomechanical aspects of two different implant-prosthetic concepts for edentulous maxillae. Int J Oral Maxillofac Implants 1995;10:188-98.  Back to cited text no. 37
    
38.
Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: A review of the literature. J Prosthet Dent 2001;85:585-98.  Back to cited text no. 38
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7]

This article has been cited by
1 Combining Intraoral and Face Scans for the Design and Fabrication of Computer-Assisted Design/Computer-Assisted Manufacturing (CAD/CAM) Polyether-Ether-Ketone (PEEK) Implant-Supported Bars for Maxillary Overdentures
Francesco Mangano,Carlo Mangano,Bidzina Margiani,Oleg Admakin
Scanning. 2019; 2019: 1
[Pubmed] | [DOI]



 

Top
Print this article  Email this article
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,273 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures

 Article Access Statistics
    Viewed1055    
    Printed20    
    Emailed0    
    PDF Downloaded124    
    Comments [Add]    
    Cited by others 1    

Recommend this journal