ORIGINAL ARTICLE
Year : 2020  |  Volume : 10  |  Issue : 2  |  Page : 177-182
Irradiance of different curing modes of common light cure devices: An in vitro study


1 Department of Restorative Dentistry, Faculty of Dentistry, King Abdulaziz University, Jeddah, Saudi Arabia
2 Al-Farabi Private College, Jeddah, Saudi Arabia

Date of Submission19-Dec-2015
Date of Acceptance07-Feb-2020
Date of Web Publication12-Mar-2020

Correspondence Address:
Dr. Hani M Nassar
Department of Restorative Dentistry, Faculty of Dentistry, King Abdulaziz University, P.O.Box 80209, Jeddah 21589.
Saudi Arabia
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jispcd.JISPCD_496_19

Rights and Permissions

   Abstract 

Aim: The aim of this study was to test the irradiance values of different curing modes of commonly available light cure devices (LCDs). Materials and Methods: An in vitro investigation was carried out to compare the irradiance output of 10 brands of LCDs available in Saudi Arabia measured using a digital radiometer. Values were recorded for three time points when applicable (0, 10, and 20s). This technique was repeated five times for each LCD. Normal, high-intensity, and soft-start modes were evaluated for all brands with the features available. Irradiance values between brands were analyzed using one-way analysis of variance followed by Bonferroni method. Changes in irradiance between different time points were analyzed using one sample t test for normal and high-intensity modes and using paired t test for soft-start mode. All comparisons were carried out at 0.05 significance level. Results: The highest values were reported for Ortholux Luminous, Elipar DeepCure-S, Elipar DeepCure, and KaVo mini-LED with values above 1000 mW/cm2. All LCDs showed values above 600 mW/cm2. Three LCDs had high-intensity mode and only one device had soft-start mode. Changes over the different time points were not statistically significant exept for soft-start mode. Conclusion: All tested LCDs had irradiance values sufficient for adequate polymerization of resin composite. Only four of these are capable of curing bulk-fill composites.


Keywords: Curing modes, irradiance, light cure


How to cite this article:
Nassar HM, Almutairi M, Makhdom A. Irradiance of different curing modes of common light cure devices: An in vitro study. J Int Soc Prevent Communit Dent 2020;10:177-82

How to cite this URL:
Nassar HM, Almutairi M, Makhdom A. Irradiance of different curing modes of common light cure devices: An in vitro study. J Int Soc Prevent Communit Dent [serial online] 2020 [cited 2020 Nov 24];10:177-82. Available from: https://www.jispcd.org/text.asp?2020/10/2/177/280523





   Introduction Top


Resin composites are used widely in the dental field, due to their excellent aesthetic and physical properties.[1],[2],[3] Resin composite formulations have advanced drastically to cope with clinical applications and patients’ demands.[4] The selection of a good light cure device (LCD) is an essential step for curing composite restoration.[3],[5],[6] The polymerization of composite particles from monomer to polymer is initiated by photoinitiator activation; light cure quality will affect the degree of polymerization reaction. The number of photons generated by LCDs affect the light intensity and subsequent polymerization reaction.[7]

The success and longevity of resin composite restorations can be affected by the quality of light generated by LCDs. A common metric to depict this is the irradiance value (measured in mW/cm2), which is described as the amount of light reaching a particular surface area of composite.[5] In general, irradiance values of <300 mW/cm2 are not recommended to cure dental composites.[8] Irradiance is affected by factors such as radiant power of the LCD and distance from composite surface.[9] The longevity of the resin composite restorations can be affected by LCD performance owing to their irradiance values.[10] This is especially true with the newly introduced bulk-fill composite materials that have different chemistry and photoinitiator formulations requiring greater irradiance values (>1000 mW/cm2) to achieve adequate polymerization.[11]

Further, the polymerization reaction of resin composites can be affected by the different modes of light curing as these modes alter the light output of the device. In the traditional mode, a burst of light with high irradiance values will lead to fast polymerization and possibly high polymerization shrinkage stress. To minimize this problem, gradual polymerization mode is found that could potentially help in reducing the formation of marginal gap.[4] High-intensity mode is sometimes used in orthodontic application where a relatively thin amount of resin is used to bond brackets. Still, irradiance values of different light-curing modes are of a great importance to the practicing dentist. Therefore, the aim of this study was to measure the irradiance values of different brands of LCDs and their curing modes.


   Materials and Methods Top


Study design

This investigation was an in vitro cross-sectional study involving 10 LCDs available in the Saudi Arabian market. The irradiance values of different curing modes were recorded.

Sampling criteria

Light irradiance of 10 light-emitting diode (LED) LCDs available in the Saudi market [Figure 1], Elipar DeepCure (3M, St. Paul, Minnesota), Elipar DeepCure-S (3M), Ortholux Luminous (3M), Bluephase N M (Ivoclar Vivadent, Schaan, Liechtenstein), VALO Cordless (Ultradent Products, South Jordon, Utah), Demi Ultra (KaVo Kerr, Orange, California), Dr’s Light Clever (Good Doctors, Incheon, Republic of Korea), KaVo mini-LED (KaVo Dental GmbH, Bismarckring, Germany), DTE-iLED (Woodpecker, Guilin, Guangxi, China), and FlashMax P3 (Ragnagade 7, Copenhagen, Denmark), was measured using a digital radiometer (Bluephase Meter II, Ivoclar Vivadent).
Figure 1: The different light cure devices included in the study. (A) Ivoclar Vivadent Bluephase N M. (B) 3M Elipar DeepCure-S. (C) Ultradent VALO Cordless. (D) 3M Ortholux Luminous. (E) Dr’s Light Clever. (F) DTE-iLED. (G) FlashMax P3. (H) 3M Elipar DeepCure-L. (I) KaVo mini-LED. (J) Kerr Demi Ultra

Click here to view


Observational parameters

The tip of the LCD was measured using the radiometer’s built in gauge and the value was plugged into the radiometer. The curing tip was laid flat on the radiometer sensor and the device was operated for 20s. Irradiance values were recorded for 0, 10, and 20s. Some devices did not support continuous curing for 20s; only values for 0 and 10s were recorded for those devices. This technique was repeated five times for each LCD and an average value was calculated. A similar approach was used for LCDs with soft-start mode. For high-intensity mode, irradiance value after 3s was reported.

Statistical analysis

Data were collected, tabulated, and analyzed using Statistical Package for the Social Sciences software program, version 17.0 (IBM, Armonk, New York). Quantitative variables were described using the mean, standard deviation, and 95% confidence interval. For irradiance data among different brands, one-way analysis of variance (ANOVA) was used to test differences in mean values followed by Bonferroni multiple comparison method to determine which devices are different from each other. Changes of irradiance values at different time points within the same brand were analyzed using one-sample t test for normal and high-intensity modes and using paired t test for soft-start mode. All comparisons were carried out at 0.05 significance level.


   Results Top


Overall, all LCDs tested registered irradiance values above the threshold of 600 mW/cm2 [Figure 2]. However, only four of these were associated with irradiance values >1000 mW/cm2 required to adequately polymerize bulk-fill formulation. Statistical testing of mean irradiance values for normal mode showed statistically significant differences among the tested brands [Table 1] and [Table 2]; P < 0.001].
Figure 2: Mean values of light intensities with normal mode of various light cure unit brands.

Click here to view
,
Table 1: One-way analysis of variance: tests between subjects effects for normal curing mode

Click here to view
,
Table 2: Bonferroni test: tests whose devices are different for normal curing mode

Click here to view


Irradiance values can be categorized into four groups: Group A with irradiance values between 600 and 800 mW/cm2 that includes Dr’s Light Clever, DTE-iLED, VALO Cordless, and Demi Ultra; Group B with values between 800 and 1000 mW/cm2 that includes Bluephase N M; Group C with values between 1000 and 1200 mW/cm2 that includes Elipar DeepCure-S, Elipar DeepCure, and KaVo mini-LED; Group D with values >1200 mW/cm2 that includes Ortholux Luminous.

Only three LCDs had a high-intensity mode [Figure 3] with FlashMax P3 showing significantly higher irradiance values as compared to DTE-iLED and VALO Cordless [Table 3] and [Table 4]; P < 0.001]. Regarding changes in irradiance, no statistical significant changes were observed for 0, 10, and 20s in all brands for normal and high-intensity modes (data not shown; P > 0.05).
Figure 3: Mean values of light intensities with high-intensity mode of various light cure unit brands.

Click here to view
,
Table 3: One-way analysis of variance: tests between subjects effects for high-intensity curing mode

Click here to view
,
Table 4: Bonferroni test: tests whose devices are different for high-intensity curing mode

Click here to view


For soft-start mode [Figure 4], only one device had this feature, which is KaVo mini-LED. Values for 0s were <300 mW/cm2 and then increased significantly at 10 and 20s (P < 0.05).
Figure 4: Mean light intensity for KaVo mini-light emitting device with soft-start mode.

Click here to view



   Discussion Top


The use of composite resin restorations is increasing nowadays due to superior esthetics and good mechanical properties.[2],[12] The irradiance values of light-curing devices are an integral part in achieving predictable composite resin polymerization and subsequent restoration life span.[13],[14],[15] Due to the introduction of new bulk-fill composite formulations, the need to ensure adequate light irradiance of curing units is even greater. The objective of the current investigation was to test the light irradiance of different light cure units in different modes.

The technology of light cure has been improved in the last decade due to foundation of high-intensity LED, quartz–tungsten–halogen (QTH) light, and plasma arc lights. These devices are manufactured to generate less heat and to cure the resin faster. Due to its advantages, LED recently is the most popular LCD as compared to halogen light.[4] Among LEDs, 10 of the most used LCDs were chosen. Some of these have high-intensity mode, which is important in orthodontic brackets placement. We used a digital radiometer because it gives reliable results and takes into consideration the tip dimeter.

In addition for testing light output for conventional composite curing, we wanted to investigate which LCDs in the market can do proper polymerization of bulk-fill composite materials. We found three devices (Ortholux Luminous, Elipar DeepCure, and Elipar DeepCure-S), which can do proper polymerization for bulk-fill composite due to their irradiance values being >1000 mW/cm2. In a recent study, only 10% of the tested 166 LCDs produced values capable of polymerizing bulk-fill formulations.[16] The bulk-fill formulation usually contains new photoinitiators that are more sensitive but still require higher irradiance values to produce adequate degree of conversion with 4mm or more increments.[17],[18] In general, bulk-fill composites require 10s of curing at >1000 mW/cm2 according to most manufacturers.[19]

Three LCUs have the high irradiance (DTE-iLED, VALO Cordless, and FlashMax P3 460), which can be used in placing orthodontic brackets. All tested LCDs can do proper polymerization of conventional composite with irradiance values of 600 mW/cm2 and above. Of course, this setup must be accompanied by using incremental technique to reduce the inherent polymerization shrinkage of resins.[20],[21] Each layer has to be light cured for at least 20s to achieve proper degree of conversion[10] and to avoid negative repercussions of inadequate polymerization that include discoloration, sensitivity, and pulpal irritation.[3],[5],[6],[22],[23] In a clinical survey, Bansal et al.[24] found that 54% of the 1000 tested devices gave output values <400 mW/cm2. On the contrary of this and in accordance with our findings, Soares et al.[10] reported that only 2 of 22 tested LCDs produced irradiance values of <400 mW/cm2. Although all the tested devices of this study were new, frequent maintenance and check is required to ensure adequate output of LCDs.

As with other in vitro studies, this investigation has some limitations. The choice of LCDs was limited to devices available in the Saudi market. Also, effect of irradiance on actual composite was not tested. Still, dentists practicing in Saudi Arabia and neighboring countries would benefit from the current instigation’s results as it has direct impact on the predictability and mechanical properties of composite restorations they place.[3],[22],[23] Further, as depicted from previous investigations, mechanical characteristics of cured composite depend on the irradiance values of LCDs and subsequent degree of conversion.[14],[15] On the basis of our results, the majority of investigated LCDs produced adequate irradiance values in their normal mode to achieve adequate polymerization of resin composite. Only three devices were able of producing values >1000 mW/cm2 and are capable of curing bulk-fill composite formulations.


   Conclusion Top


Using light-curing units with adequate irradiance and exposing each composite resin increment for the recommended curing time are important factors to achieve restorations with predictable properties and long service life.

Acknowledgments

None.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

Author contributions

Study conception (HN), data collection (AM, MA), data acquisition and analysis (AM, MA), data interpretation (HN, AM, MA), manuscript writing (HN, AM, MA).

Ethical policy and institutional review board statement

This project is exempted from ethical approval due to lack of human and/or animal samples. All the procedures have been performed as per the ethical guidelines laid down by Declaration of Helsinki.

Data availability statement

The data that support the study results are available from the author (Dr. Albaraa Makhdom, e-mail: [email protected]) on request.



 
   References Top

1.
Demarco FF, Corrêa MB, Cenci MS, Moraes RR, Opdam NJ. Longevity of posterior composite restorations: Not only a matter of materials. Dent Mater 2012;28:87-101.  Back to cited text no. 1
    
2.
Jadhav S, Hegde V, Aher G, Fajandar N. Influence of light curing units on failure of directcomposite restorations. J Conserv Dent 2011;14:225-7.  Back to cited text no. 2
[PUBMED]  [Full text]  
3.
Hao X, Luo M, Wu J, Zhu S. A survey of power density of light-curing units used in private dental offices in Changchun city, China. Lasers Med Sci 2015;30:493-7.  Back to cited text no. 3
    
4.
Poggio C, Lombardini M, Gaviati S, Chiesa M. Evaluation of Vickers hardness and depth of cure of six composite resins photo-activated with different polymerization modes. J Conserv Dent 2012;15:237-41.  Back to cited text no. 4
  [Full text]  
5.
Shimokawa CA, Harlow JE, Turbino ML, Price RB. Ability of four dental radiometers to measure the light output from nine curing lights. J Dent 2016;54:48-55.  Back to cited text no. 5
    
6.
Maghaireh GA, Alzraikat H, Taha NA. Assessing the irradiance delivered from light-curing units in private dental offices in Jordan. J Am Dent Assoc 2013;144:922-7.  Back to cited text no. 6
    
7.
Son SA, Park JK, Jung KH, Ko CC, Jeong CM, Kwon YH. Effect of 457nm diode-pumped solid state laser on the polymerization composite resins: Microhardness, cross-link density, and polymerization shrinkage. Photomed Laser Surg 2015;33:3-8.  Back to cited text no. 7
    
8.
Bansal R, Hora BS, Kumar A, Bansal M, Gupta C, Singla S. Are we doing justice? A clinical survey of the output intensity of light curing units in dental offices. Int J Dent Sci 2012;4:9-11.  Back to cited text no. 8
    
9.
Price RB, Ferracane JL, Shortall AC. Light-curing units: A review of what we need to know. J Dent Res 2015;94:1179-86.  Back to cited text no. 9
    
10.
Soares CJ, Rodrigues MP, Oliveira LRS, Braga SSL, Barcelos LM, Silva GRD, et al. An evaluation of the light output from 22 contemporary light curing units. Braz Dent J 2017;28:362-71.  Back to cited text no. 10
    
11.
El-Damanhoury H, Platt J. Polymerization shrinkage stress kinetics and related properties of bulk-fill resin composites. Oper Dent 2014;39:374-82.  Back to cited text no. 11
    
12.
Rueggeberg FA, Giannini M, Arrais CAG, Price RBT. Light curing in dentistry and clinical implications: A literature review. Braz Oral Res 2017;31:e61.  Back to cited text no. 12
    
13.
Ward JD, Wolf BJ, Leite LP, Zhou J. Clinical effect of reducing curing times with high-intensity LED lights. Angle Orthod 2015;85:1064-9.  Back to cited text no. 13
    
14.
AlShaafi MM, Harlow JE, Price HL, Rueggeberg FA, Labrie D, AlQahtani MQ, et al. Emission characteristics and effect of battery drain in “budget” curing lights. Oper Dent 2016;41:397-408.  Back to cited text no. 14
    
15.
Shortall A, El-Mahy W, Stewardson D, Addison O, Palin W. Initial fracture resistance and curing temperature rise of ten contemporary resin-based composites with increasing radiant exposure. J Dent 2013;41:455-63.  Back to cited text no. 15
    
16.
Nassar HM, Ajaj R, Hasanain F. Efficiency of light curing units in a government dental school. J Oral Sci 2018;60:142-6.  Back to cited text no. 16
    
17.
Arikawa H, Shinohara N, Takahashi H, Kanie T, Fujii K, Ban S. Light transmittance characteristics and refractive indices of light-activated pit and fissure sealants. Dent Mater J 2010;29:89-96.  Back to cited text no. 17
    
18.
Lima RBW, Troconis CCM, Moreno MBP, Murillo-Gómez F, De Goes MF. Depth of cure of bulk fill resin composites: A systematic review. J Esthet Restor Dent 2018;30:492-501.  Back to cited text no. 18
    
19.
Menees TS, Lin CP, Kojic DD, Burgess JO, Lawson NC. Depth of cure of bulk fill composites with monowave and polywave curing lights. Am J Dent 2015;28:357-61.  Back to cited text no. 19
    
20.
Shimokawa CA, Turbino ML, Harlow JE, Price HL, Price RB. Light output from six battery operated dental curing lights. Mater Sci Eng C Mater Biol Appl 2016;69:1036-42.  Back to cited text no. 20
    
21.
Morimoto S, Zanini RA, Meira JB, Agra CM, Calheiros FC, Nagase DY. Influence of physical assessment of different light-curing units on irradiance and composite microhardness top/bottom ratio. Odontology 2016;104:298-304.  Back to cited text no. 21
    
22.
Ferracane JL. Resin-based composite performance: Are there some things we can’t predict? Dent Mater 2013;29:51-8.  Back to cited text no. 22
    
23.
Heo YJ, Lee GH, Park JK, Ro JH, García-Godoy F, Kim HI, et al. Effect of energy density on low-shrinkage composite resins: Diode-pumped solid state laser versus quartz-tungsten-halogen light-curing unit. Photomed Laser Surg 2013;31:28-35.  Back to cited text no. 23
    
24.
Bansal R, Bansal M, Shilpa W, Loveena B, Karanvir S, Ridhi A.Assessment of efficacy and maintenance of light-curing units in dental offices across Punjab: A clinical survey. Indian J Dent Sci 2019;11:42-5.  Back to cited text no. 24
  [Full text]  


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
Print this article  Email this article
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,014 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed524    
    Printed26    
    Emailed0    
    PDF Downloaded93    
    Comments [Add]    

Recommend this journal